AUTO POWER OFF CONFIGURATION

The default setting for sleep mode is 120 seconds. The sleep mode can be

configured to 120, 240 seconds or OFF via keyboard. If there is no activity, the scale

will turn itself off after 120 seconds.

The Auto Power Off feature can be changed, when needed, by the following

instructions:

Function

Scale Response

With the scale turned on,

remove all weight from the scale

Press and hold the “TARE” key

for 3 seconds . . . . . . . . . . . . . . . . . . . . . . . . . “TARE”, “SET”, “Unit”

Press the “UNITS” key to scroll

to the next selection . . . . . . . . . . . . . . . . . . . . .“A.OFF”

Press “TARE” key to enter into the

AUTO POWER OFF configuration menu . . . . . .”120” (120 seconds)

Press the “TARE” key to accept this setting or the “UNITS” key to scroll

between “240” (240 seconds) or “OFF.

After selecting your desired setting

and pressing the “TARE” key to accept. . . . . . “A.OFF”

Press the “UNITS” key twice to scroll to . . . . . .“END”

Press the “TARE” key to exit the setting and return the scale to “0.0” weight.

5

SCALE CALIBRATION INSTRUCTIONS

Periodically your scale may require calibration to ensure proper weighing.

Calibration may be required if the scale has been improperly used by mishandling

or dropping the scale. You will need an accurate 50 or 200 pound weight to

calibrate the scale. If calibration is performed with a weight other than 50

or 200 pounds, it will cause your scale to weigh inaccurately.

IMPORTANT…Do not Enter into Calibration Mode without an accurate weight.

Scale calibration is accomplished, when needed, by the following instructions:

Function

Scale Response

1. Make sure the scale is turned off

2. Remove all weight from the scale

3. Press and hold the “ZERO” key and

press the “UNITS” key three times . . . . . . . . ... .”CAL”

4. Press the “UNITS” once to accept and

enter into calibration mode. . . . . . . . . . . . . . . . . ”1000 – 4000”

5. Press the “TARE” key once

to zero the scale. . . . . . . . . . . . . . . . . . . . . . . . ..“0”

6. Place “XX” (see Table 1)

weight on the scale . . . . . . . . . . . . . . . . . . . . . .. See Table 2

7. Press “TARE” key once

to accept the calibration weight . . . . . . . . . . . . . . . . . . . .“CAL”

8. The display prompt “CAL” for 2-3 seconds, then exit the

calibration mode and return to normal weighing mode. If the

calibration weight is left on the scale, a weight will be displayed.

Table 1

Table 2

Model

Calibration

Weight

(LB)

Calibration

Weight

(KG)

Span Range

(LB)

Span Range

(KG)

GP100

50 LBS

20 KG

900 - 1600

800 – 1400

GP250

200 LBS

100 KG

3600 – 6400

4000 – 7000

6