Salter Brecknell PS500 User Manual

Page 17

Advertising
background image

.Key Definition summary:

KEY MODE

DEFINITION

Normal weighing mode

Enter or exit HOLD mode; output the
data as per P4,P5,P6 setting

Setup mode/Calibration

mode

Shift the flashed position from right to
left

HOLD

/PRINT

Displaying A/D code or input

voltage mode

No Function

Normal weighing mode

Choose weight units, refer P8, P9, P10,
P11 setting and Table3,Table4.

Setup mode/Calibration

mode

Change the digit on flashed position and
click this button to add 1.

UNIT

Displaying A/D code or input

voltage mode

Choose the weight inner code or input
working voltage to be displayed.

Normal weighing mode Tare the weight

Setup mode/Calibration

mode

Confirm the displayed parameters or
input data, and go to next step

TARE

Displaying A/D code or input

voltage mode

Choose filtered or un-filtered weight
A/D data

ON/OFF/

ZERO

Normal weighing mode Zero function or Power off the scale.

26

Setup mode/Calibration

mode/Displaying A/D code

or input voltage mode

Exit and back to normal work mode

ON/OFF/

ZERO

+

HOLD

/PRINT

Normal weighing mode

(more than 3s)

Go to Show A/D code or input working
voltage of indicator mode

ON/OFF/

ZERO

+

UNIT

Normal weighing mode

(more than 3s)

Enter setup mode

ON/OFF/

ZERO

+

TARE

Normal weighing mode

(more than 3s)

Enter calibration mode when

.Parameters setting summary: SEE SPECS

Para-

meter

x/xy

Factory

Set

Setting

P1.xy

00-15

05

Auto-off time: no auto-off;
01-15 minutes auto-off time

P2.x

0, 1, 2

2

0=only Hold function; 1=only Print function;
2=Print and Hold function,

P3.xy

0,1,2,

3-50

2

0=no hold function; 1=hold lager weight reading;
2=auto release hold function when weight is below
10d and auto-hold new stable weight (more than
10d); 3-50=unchangeable reading when the variety
is within +3~50d

27