Salter Brecknell 400ES User Manual

Page 30

Advertising
background image

Page D - 1

APPENDIX D: DISPLAYED ERROR CODES

CODE

MODE

MEANING / POSSIBLE SOLUTION



Normal Operating
Mode

Gross Overload. A weight greater than the rated capacity has been
applied to the scale. Remove the weight from the platter or try re-
calibrating the scale. Otherwise, check for a bad load cell connection or
possible load cell damage due to overloading.

Err 0

Span Calibration
Mode (F17)

Keyed-in weight value is larger than full scale capacity. Use a smaller test
weight or check keyed-in value.

Err 1

Span Calibration
Mode (F17)

Keyed-in weight value is less than 1% of full scale capacity. Use a larger
test weight or check keyed-in value.

Err 2

Span Calibration
Mode (F17)

There is not enough load cell signal to produce the internal counts
necessary to properly calibrate the scale. First check all load connections.
Use F16 mode to view internal counts. See Appendix C for more
information.

Err 3

All Modes

Non-volatile memory read error. One or more setup parameters have
been lost.

Err 4

All Modes

Non-volatile memory write error. Indicator needs service.

Err 5

Key-in Span
Calibration Mode
(F20)

You have attempted to enter a zero value for C1. Enter a known
calibration value greater than zero.

Err 7

Initialization

No reading from the ADC. Make sure there is a load cell(s) connected to
the indicator at start-up.

Err 9

Normal Operating
Mode

Span calibration value has been lost. Re-calibrate the scale.

t Err 1

Normal Operating
Mode

You have attempted to key-in a tare weight that is greater than the scale’s
maximum capacity. For example, on a 6,000 lb scale, you cannot enter,
say, 7,000 lb.

t Err 6

Normal Operating
Mode

You have attempted to key-in a tare weight whose resolution is higher
than the resolution of the scale For example, on a 100 x 0.02 lb scale, you
cannot enter, say, 10.01 lb since the resolution is 0.02 lb.