Salter Brecknell HS-300 User Manual

Page 16

Advertising
background image

26

ON/OFF/

ZERO

+

UNIT

Normal weighing mode

(more than 3s)

Enter setup mode when sealed
calibration switch is on?

ON/OFF/

ZERO

+

TARE

Normal weighing mode

(more than 3s)

Enter calibration mode when sealed
calibration switch is on?

.

Parameters setting summary:

Para-

meter

x/xy

Factory

Set

Setting

Final

Set

P1.xy

00-15

05

Auto-off time: no auto-off; 01-15
minutes auto-off time

P2.x

0, 1, 2

2

0=only Hold function; 1=only Print
function; 2=Print and Hold function,

P3.xy

0,1,2,

3-50

2

0=no hold function; 1=hold lager weight
reading; 2=auto release hold function
when weight is below 10d and auto-hold
new stable weight (more than 10d);
3-50=unchangeable reading when the
variety is within +3~50d

P4.x

0,1,2,3,4,

5,6,7

2

0=no RS232 function; 1=output display
data when PRINT pressed; 2=output
gross, tare and net weight when PRINT
pressed; 3=continuously output display
data; 4= continuously output gross, tare
and net weight; 5=output display data
one time when scale is stable; 6=output
gross, tare and net weight one time when
scale become stable; 7=Bio-RS232
compatible (8 data) to NCI-SP1;

P5.x

0, 1, 2, 3,

4

3

Baud rate for RS232: 0=1200bps,
1=2400bps, 2=4800bps, 3=9600bps,
4=19200bps

27

P6.x

0, 1, 2

0

RS232 format: 0=8N1, 1=7O1, 2=7E1

P7.xy

00-31

13

Resolution select: 500,600,750,800,
1000,1200, 1500,2000,2400,2500, 3000,
3500, 4000, 5000, 6000, 7000, 7500,
8000, 10000, 12000, 15000, 20000,
25000, 30000, 35000, 40000, 50000,
60000,70000,75000,80000,100000

P8.x

0,1, 2

0

Division select: 0=1, 1=2, 2=5

P9.x

0,1,2,3,4,

5

0

Decimal point in calibration: 0= x1, 1=
x0.1, 2= x0.01; 3= x0.001; 4= x0.0001;
5= x10

P10.x

0, 1

1

Calibration unit: 0=kg, 1=lb

P11.x

0, 1, 2, 3,

4, 5, 6

6

Weighing units enable: 0=only kg;
1=only lb; 2=only lb:oz; 3=kg or lb;
4=kg or lb:oz; 5=lb or lb:oz; 6=kg, lb, or
lb:oz

P12.x

0, 1, 2, 3,
4, 5, 6, 7,

3

Power-on zero-point range:
0=calibration zero -point +1%FS;
1=calibration zero -point +2%FS;
2=calibration zero-point +5%FS;
3=calibration zero-point +10%FS;
4=calibration zero-point +20%FS;
5=calibration zero-point +50%FS;
6=calibration zero-point +100%FS;
7=No limitation

P13.x 0,1….14

4

Zero range for ZERO button:
0= Power-on zero-point ±1%FS;
1= Power-on zero-point ±2FS;
2= Power-on zero-point ±3%FS;
3= Power-on zero-point ±4%FS;
4= Power-on zero-point ±5%FS;
5= Power-on zero-point ±10%FS;
6= Power-on zero-point ±20%FS;
7= No limitation
8= Power-on zero-point +1%FS
9= Power-on zero-point +2%FS
10=Power-on zero-point +3%FS