Technical Specifications

Scale Indicator
Input signal range:

0mV ~ +30mV

Sensitivity:

>0.2uV/grad

Internal Resolution:

Approximately 520,000 counts

Display Resolution:

Can be selected between 500-100,000

System Linearity:

Within 0.01% of FS

Loadcell excitation Voltage:

+5 V

DC

(MAX current: 85mA)

Loadcell
Sensitivity:

0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad.)

Input Resistor:

≥60 Ω

Output Resistor:

<10 KΩ

Temperature
Operation:

5

o

C - 35

o

C

Storage:

10

o

C - 70

o

C

Humidity: ≤95%RH (no condensation)

Power

Battery:

6V4AH lead acid battery, 30 hrs continuous use

When the voltage of battery is below 5.7V, the “Low Bat” annunciator will be lighted, plug in AC adapter to charge
the battery. When “Lo.bAt” and actual weight is displayed alternately, this indicates the voltage of battery is below
5.5V and the scale will be turned off in two minutes automatically.

AC Adapter:

10.5-12V

DC

600mA, with central positive, 5.7 VDC -9 VDC

Error Messages

Error Message

Definition

Required
Solution/Troubleshooting


0¯ ¯ ¯ ¯:

Weight above range for calibrated
zero point.

Remove load before zeroing
Or
Recalibrate the scale.



0_____:

Weight below range for calibrated
zero point.

Remove load before zeroing
Or
Recalibrate the scale.


_____:

Indicates an under range condition Remove all loads and zero the

scale.


¯ ¯ ¯ ¯:

Capacity exceeded

Remove the load and try again. If
the load is great than 6 kg a scale
with a larger capacity is required.

CAL-Er: Calibration

error

Restart

calibration

Lo.bAt:

Low Battery

Recharge the battery. Upon initial
use, it is recommended to charge
battery for 8 hours prior to use.

EEP.E0

EEPROM can’t be accessed

Replace S100 Indicator

EEP.E1

Configuration settings have
changed and not been stored

Calibrate the scale to store
settings

EEP.E2

P7-P9 settings exceed scale’s
normal range

Reconfigure P7-P9 setting