67XXU USB/Serial Scale Service Manual

LINE

CAL.P0

Line calibration point0

Do zero point calibration. This point
can't be omitted.

CAL.P1

Line calibration point1

First weight point calibration. This
point can't be omitted and standard
weight must be over 10%FS.

END.Y

YES

End calibration?

YES=go to CAL.END to end
NO=go to do next point calibration

NO

CAL.P2

Line calibration point2

Second weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P1. This point can be omitted.

END.Y

YES

End calibration?

YES=go to CAL.END to end
NO=go to do next point calibration

NO

CAL.P3

Line calibration point3:

Third weight point calibration.
Standard weight must be over
10%FS and be larger than it in
CAL.P2, this point can be omitted.

geo

code

00-70

Select geograhical position

Code 00-70

gravt

9.76183-

9.99999

Input gravity of user location

input

Input or view calibration
parameter values

Only used to copy calibration
parameters from one scale to a new
scale.

CAL

Submenu1

Submenu2

Option Parameter Description

Comment