CALIBRATION

CHAPTER 5

6.

Enter the following code into the display using the Cursor Keys:

5222


( Left (

←), Right (→) to select digit ; Up (↑), Down (↓) = increment/decrement value ; Enter (↵) to accept )


Calibration CW:

7. Select

Calibrate

CW. Press ENTER (

↵).

A warning is displayed.

Select ‘

Yes

’. Press

Enter

(

↵) .

8. Using the Up (

↑) or Down (↓) Cursor keys, enter “

1

” for a

single transducer,

or enter “

4

” for the

4-in-1

transducer.

9.

Press ENTER (

↵). The display reads:

Select transducer by applied torque on it, LED will lit.

5-8