To determine the additional amplifier overhead required to handle 80% amplitude

modulation, multiply the result by 3.24 (1.8

2

). Actual transmitted field strength

should be verified using an ETS-Lindgren Model HI-6005 Field Probe, or

equivalent. The forward power data graphs on page 39 and page 40 show power

requirements for the lower frequencies at three meters based on measurements

using a field probe on an Open Area Test Site (OATS) over both conducting

ground and a (2.4m)

2

ferrite absorber field over conducting ground. For

IEC 61000-4-3 type testing, the antenna tip can be placed at any distance

between one and three meters from the Equipment Under Test (EUT) as long as

the front face plane is illuminated according to the -0, +6 dB uniform field

specification. In general, closer distances require less power to create a given

field strength.