ETS-Lindgren 2005 Single Axis Positioner User Manual

Page 23

Advertising
background image

Assembly and Installation

| 23

Filter Installation

The filter is equipped with a 1 in. inside / 1.3 in. outside diameter (2.5cm

inside/3.3cm outside) standard pipe penetration conduit kit. This includes pipe,

conduit nut, brass nut, washer, circular EMI gasket and plastic cap. A 1.35 in.

(3.429cm) hole must be made in the shield of the room in order to accommodate

the conduit penetration of the filter. Drawings are included for reference during

installation.

Open the gasketed compartment of the filter, being careful not to

damage the EMI gasket that sits below the cover.

Remove the plastic cap, brass nut and washer from the conduit

penetration on the outside of the filter.

Clean the shield using a scouring pad (provided by the customer)

suitable for cleaning stainless steel surfaces ensuring no glue residue

is present.

Failure to use an appropriate scouring pad to

remove all residue could result in reduction of

shielding effectiveness.