ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual

Page 26

Advertising
background image

Page — 18

HI

n

1501 Manual

Figure 3

Oven Scanning

switch to the “SLOW” position. Return to the

spot where you noticed the highest reading

before.

7.

Hold the probe in place over the spot for at least

five seconds and watch for the highest needle

indication. The needle will move slower and not

go quite as high as it did in the “FAST” position.

The reading you obtain is the maximum leakage

of your oven.

8.

Check the zero after each measurement. If a

shift in the zero is observed, the meter will be

rezeroed and the measurement repeated.

9.

Turn the selector switch to OFF when the meter

is not in use.