Operation

MODEL 2075 MINIMAST™

18

© EMC TEST SYSTEMS, L.P. – MARCH 2002

REV D – PN 399235

OPERATION


Please refer to the Model 2090 Positioning Controller manual if
you are unfamiliar with the operation of the unit. A Model 2090
manual is included with each 2090 shipment and it also available
for download from our website.

EDITING MODEL 2090 POSITIONING
CONTROLLER CONFIGURATION
PARAMETERS


To edit a configuration parameter, press the PARAM key to
display the current parameter. Pressing the PARAM key repeatedly
will scroll down through the parameter list, showing each
parameter in turn. While viewing a parameter, the STEP keys
(INC/DEC) may be used to scroll up or down the parameter list.
This reduces the effort necessary to scan through a long parameter
list using the PARAM key. Pressing any of the LIMIT/POSITION
selection keys will return the display to that selection. Pressing
any of the remaining motion keys will return the display to the
current position and execute that motion. Pressing the PARAM
key again will return to the last displayed parameter in the list,
allowing easy transition between parameter adjustment and device
operation.

Once the desired limit, position or parameter is visible in the
display window, pressing INCRM, DECRM, or ENTER will
toggle into edit mode. The lowest adjustable digit will flash on
and off. Pressing the LOCAL key for that device will switch the
flashing digit to the next higher digit. In this way, it is possible to
rapidly adjust any digit of a multi-digit parameter or limit.

TOWER ENCODER CALIBRATION


The display symbol

C refers to the encoder calibration parameter.

This setting is used to convert the encoder count values returned
from a motor base into the corresponding centimeter or degree
position reading. For towers, the number represents the number of