APPENDIX A
Command Set (RS-232 Serial Port)
Function
Command Range
Default
Set alarm trigger point
T#
0-9999
-
Set filter number
F#
1-4
1
Set range
R#
1-4
1
Set peak hold (1=ON)
P#
0-1
0
Zero meter
Z
-
-
Initialize system (Reset)
I
-
-
Set 3 or 4 digit display
D#
3-4
3
Set print mode
M#
0-9
2
Set baud rate
B#
12-96
12
Set auto-range (1=ON)
A#
0-1
0
Set quiet mode (1=ON)
Q#
0-1
1
Set command echo (1=ON) E#
0-1
0
Run self test
ST
-
-
Set status line
S#
0-8
-
0
Send system status
1
Send alarm trigger point
2
Send range number
3
Send filter number and filter spec.
4
Send bias voltage
5
Send offset (in milliwatts)
6
Send stim value
7
Send power supply voltages
8
Send current RF reading