Display of the results

The procedure having run normally, the module having detected the
beginning and the end of the salt cloud, it will now, upon pressing
touch

, display the data as shown here at the right. If at the

moment of the display END OF CLOUD the measurement of salinity
has reached its initial value, we may take it for granted that all the
injected salt has passed though the gauging area. We may then
graphically check the quality of the salt dilution on the PC with help
of the SALINOMADD software and eventually correct the points of
beginning and end of salt cloud.

It may happen that the SalinoMADD does not detect the end of the
cloud and thus the operator should interrupt manually the
procedure. In this case the display says MANUAL STOP, ESTIMATED
FLOW. The estimated value should in fact be correct because the

beginning of the cloud had been detected correctly by the module and
the end by the operator.

When the memory of a site is full, the device displays MEMORY
EXHAUSTED. Various raisons can be the cause:

• The length of the salt cloud has exhausted the memory size.
• The SalinoMADD has not detected the end of the salt cloud and

the operator did not stop it.

• The end of the cloud has detected, but the 100 preliminary

measurements had not been recorded.


In any case, check the gauging curve in the graph and decide if the gauging is good or not.

End of acquisition due to errors

If the SalinoMADD has not been in position to collect a sufficient
amount of parameters for the calculation of water flow, it will explain
it on the screen.
In this case, it may be necessary to repeat the gauging procedure.
The operator, who is equipped with a laptop with SALINOMADD
software, can check the gauging curve on site and therefore leaves
the site with the assurance of perfect gauging results.