the applicable requirements of relevant

electromedical standards, including EN

60601-1-2:2007.

The my5™ model must emit

electromagnetic energy in order to

perform its intended function. Nearby

electronic equipment may be affected

by this emission. Similarly the my5™

model may itself be affected by

electromagnetic emissions from other

equipment in the vicinity.

The my5™ model is intended for use

in the electromagnetic environment

specified below. The customer or user of

the my5™ model should assure that it is

used in such an environment:

– Floors should be wood, concrete or

ceramic tile. If the floor is covered with

synthetic material, the relative humidity

in the room should be at least 30%.

– Mains power quality should be that of a

typical domestic, commercial or hospital

environment as appropriate.

– The my5™ model should not be used

during power mains interruptions.

– Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical commercial

or hospital environment.

The my5™ model is intended for

use in an electronic environment in

which radiated radio-frequency (RF)

disturbances are controlled.

The customer or user of the my5™

model can help prevent electromagnetic

interference by maintaining a minimum

distance between portable and mobile RF

communications equipment (transmitters)

and the my5™ model as recommended

below, according to the maximum output

power of the communications equipment:

The adequate separation distance

to provide base immunity to RF

disturbances is 1.0 metres.

Retain this user manual for your future

reference.