0 troubleshooting and testing – Rice Lake Bariatric Handrail (250-10-2) - Rice Lake Digital Handrail Scale Operation Instructions User Manual

Page 15

Advertising
background image

Digital Handrail Scale Operation Manual - Troubleshooting and Testing

11

6.0

Troubleshooting and Testing

Refer to the following to check and correct any failure before contacting service personnel.

Table 6-1. Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adapter.

Questionable weight or the scale does not
zero

External object is interfering with the
scale.

Remove the interfering object from the scale.

Display did not show 0.0 before weighing.

Help the patient off the scale, zero the scale and begin
the weighing process again.

Scale is not placed on a level floor.

Ensure the scale is level and begin the weighing
process again.

Scale is out of calibration.

Check the weight with a known weight value.

The display shows E messages as detailed below.

E06

Identifier

ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale.

E4L

BAT

Battery low but still useable

One bar left on the

indicator display.

E4U

Battery low and unstable

no bars left on the

indicator display.

E11

CAL

Calibration Error

recalibrate the scale again.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check the
cables and mechanical connections. If the problem
persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate.

Check the load cells’ mechanical surroundings and see
that nothing touches them and that the cables are
properly welded.

SAT

Damaged load cell cable

Replace load cell cable.

Advertising