Bariatric Handrail Scale Operation Instructions - Troubleshooting

11

5.0

Troubleshooting

Refer to the following instructions to check and correct any failure before contacting service personnel.

Table 5-1. Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adaptor.

Questionable weight or the scale does
not zero

External object is interfering with the
scale

Remove the interfering object from the
scale.

Display did not show 0.0 before
weighing

Help the patient off the scale, zero the
scale and begin the weighing process
again.

Scale is not placed on a level floor

Ensure the scale is level using the spirit
level on the platform and begin the
weighing process again.

Scale is out of calibration

Check the weight with a known weight
value.

Improper tare

Place the patient on the scale. Press
REWEIGH. Once the weight of the item
is displayed, press TARE. Place the
patient back on the scale. Press the
REWEIGH button again.

Weighing is performed but the display
shows WEIGH and REWEIGH every few
seconds; the weighing process takes too
long and no weight is displayed

The patient is not sitting still

Ask the patient to be still.

The display shows a STOP message

The load on the scale exceeds the
capacity of the scale (220 kg)

Remove the excess weight and use the
scale according to manufacture’s specs.

The display shows LO Bat message

The battery is low

Recharge the battery.

The display shows Err message as
detailed in the table below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly.
Check the cables and mechanical
connections. If the problem persists,
replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical
surroundings and see that nothing
touches them and that the cables are
properly welded.