4 attaching the feet to the scale – Rice Lake Dual Ramp Wheelchair Scale (350-10-3) - RL350-10 Technical & Operating Instructions User Manual

Page 9

Advertising
background image

Wheelchair Scale Technical Manual - Scale Assembly

5

3.4

Attaching the Feet to the Scale

The Rice Lake Digital Wheelchair Scale comes with four feet that must be attached to the scale base for proper

operation. The four feet are located in the parts box in the original shipping box. Use the following steps to attach

the feet to the scale.

1. Screw the four round feet clockwise into the scale base as shown in Figure 3-4.

Figure 3-4. Screw Feet Into the Scale Platform

2. There must be adequate clearance between the scale base and the floor so screw each foot out

counterclockwise two full turns. This will ensure that there is enough clearance between the scale base

and the floor.

3. Gently set the scale base down to the floor. There should be minimal clearance between the scale base

and the floor without having the scale base actually touching the floor. Figure 3-5 shows the width of a

finger being slid under the scale base and the floor.

Note

By not having clearance around the scale base will create inaccurate weighments.

Figure 3-5. Have a Minimal Amount of Clearance Between the Scale Base and the Floor

4. It is also important to make sure that the scale is completely level. Gently press down on all corners of

the scale base to ensure that there are no high spots or rocking of the scale base.

Note

An unlevel base will produce inaccurate weight readings.

Advertising