0 rs-232 communication, 1 pushbutton keypad print, 2 usb connection – Rice Lake Dual Ramp Wheelchair Platform Scale with Seat (350-10-4) User Manual

Page 15

Advertising
background image

Wheelchair Scale with Seat Operation Manual - RS-232 Communication

11

5.0

RS-232 Communication

The scale comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a
computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing
Systems. Figure 5-1 on page 11 shows where the RS-232 connection is.

The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity
and no handshaking.

There are three methods of communication:

Pushbutton keypad print

Escape protocol

Maintenance protocol

5.1

Pushbutton Keypad Print

With a stable, in-range weight, press and hold the

LB/KG/Print

key for at least three seconds, or until the scale

displays

PRINT

. Note that if the scale does not beep after five seconds, then release the button as the weight was

either in motion, or out of range.

If displaying weight and not BMI, the scale will send out the following 21 character string:

xxxxxxxxx<SP>uu<SP>mmmmm<SP><CR><LF>

Where:

xxxxxxxxx is the weight with decimal point and “ - “ sign, if negative, uu is the unit (lb or kg).
mmmmm is the mode (gross or net)

Example:

60.1 KG= <PATIENT><SP><WEIGHT><SP>-60.1<SP>KG<SP><CR><LF>

In BMI mode (displaying the BMI value), the scale will send out the following data:

PATIENT WEIGHT 60.1 KG
PATIENT HEIGHT 170.0 CM
PATIENT BMI 20.8

5.2

USB Connection

The Rice Lake Digital Wheelchair Scale with seat has the capability of connecting to a PC using a USB
connection and a USB cable (not included). That connection location is shown in Figure 5-1.

Figure 5-1. USB Connection Port

USB Connection

Load Cell Connection
Under This Cover

RS-232 Connection