Page 3 of 7

5)

The MSI-9000 CellScale next prompts for the capacity. With the numeric keys, enter the
capacity of the scale. Type capacity value and press ENTER.





The capacity represents the highest operating value of the crane scale. Do not enter a higher number than the
scale is rated for. The actual overload point is a few scale divisions above the entered capacity.


6)

The MSI-9000 CellScale next prompts for the count by (d). Press letter value
corresponding to desired value






This list is for a type ‘A’ scale with up to 10000 counts resolution. This is the standard default. MSI recommends
all MSI-6260CS Series crane scales calibrated at 5000 divisions or less. Too many divisions will compromise
stability and settling time. Use the largest division size suitable for your application.


7)

Unload the scale, make sure it is stable, and press Y. The scale performs its zeroing
step, removing all offset. This process takes several seconds.






















The number presented at the bottom of the screen is direct

A/D counts

. This is for troubleshooting only. This

number will never be steady as it represents 24 bits. If the deviation is less than ±250 counts the load is stable
enough for a 5000 count scale. The calibration step needs the scale platform to remain steady and unloaded
during this step
.





***** Select a countby *****

(a)

1.

Kilograms

(b)

2.

Kilograms

(c)

5.

Kilograms

(d)

10.

Kilograms

(e)

20.

Kilograms

(f)

50.

Kilograms

(g)

100.

Kilograms

***** Channel Calibration *****

Calibrating Ch # 1-1, named:

...............

Precision in Kilograms :

10000 x 2.

Test weight # 0 in Kilograms :

0.



Unload the scale.

Press (Y)

when scale is unloaded and steady

Press (B)

to backup

Press (ESC) to exit calibration

Raw A/D counts: 8403115

Type in the new capacity:

10000

kg