Hold Function

1. Press

HOLD

, when weight stabilizes display will lock and arrow will be steady.

2. Press

HOLD

to release locked weight.

Adjusting Auto-Off

1. Press

o

POWER ZERO

to power on.

2. Press and hold

UNITS

Units

. When AOFF displays, release.

3. Press

HOLD

to select auto-off time (OFF, 120, 180, 240, 300 seconds).

4. Press

UNITS

Units

to return normal weighing mode.

Calibration

Calibrate using the following process:

1. Press and hold

o

POWER ZERO

.

2. Press

UNITS

Units

three times, ZERO COUNT displays.

3. Press

o

POWER ZERO

, zero displays.

4. Press

UNITS

Units

to select calibration unit.

5. Load appropriate amount of weight on scale (See Table 3). Display shows span count.

6. Press

o

POWER ZERO

, CAL displays, then the weight.

7. Remove the weight, re-power the scale and calibration is complete.

Model

kg

lb

Table 3. Calibration Weights

OS10

10kg

20lb

OS25

20kg

40lb

OS45

20kg

40lb