Rice Lake Wheelchair Scale Operation Manual - Scale Setup

5

2. Carefully unwrap the load cell cable located on the underside of the scale. It will run out the underside of

the scale.

Figure 3-4. Unwrap the Cable from the Underside of the Scale

3. Gently set the scale base down to the floor. There should be minimal clearance between the scale base

and the floor without having the scale base actually touching the floor.

Note

Note

Not having clearance around the scale base will create inaccurate weighments.

4. It’s also important to make sure that the scale is completely level. Gently press down on all corners of the

scale base to ensure that there are no high spots or rocking of the scale base.

An un-level base will produce inaccurate weight readings.

3.5

Cable Connections

Ten feet of 4-wire cable to connect the scale to the weight indicator is supplied with each scale.

Note

The cable is pre-installed from the factory.

The cable must be routed to the indicator in a manner that will protect the cable from damage.

Cabling should exit the underside of the scale on the portable wheel side of the scale and not the ramp side(s).

When planning cable routing, leave a loose coil of excess cable under the scale to facilitate future lifting of the

scale for servicing or cleaning.

C a b l e s h o u l d
exit underside
of the scale on
t h e w h e e l e d
side of scale.

Figure 3-5. Indicator Cable Exiting the Side of the Scale Platform

5. When the interface cable is protected and in its final position, complete connections to the indicator as

described in Section 4.0.