2 offset, 3 variable, 1 coarse control range – Rice Lake Simulators User Manual

Page 11: 2 fine variable setting, 2 offset 3.3 variable, Coarse control range, Fine variable setting

Advertising
background image

Ranger 7 Simulator Operation

7

3. Calculate the required wattage using

((ExcV x ExcV)/R) x 1.5

. Use the next highest wattage value.

In a 16 load cell example, the 22 ohm resistor will have to be rated at least 6.8W. 7W is a common size.

If the excitation voltage is higher, a higher wattage resistor will be required.

Note

To achieve highest performance, low thermal EMF wiring and components are required. A manufacturer of the
binding posts is Low Thermal Electronics, they can provide low thermal spade lugs to mate to the binding
posts.

3.2

Offset

Offset will add or subtract up to ±25µV/V (±0.025mV/V).

1. Press the

Offset

switch to enable the function.

2. Rotate the knob to the desired setting.
3. Press the

Offset

switch again to disable the function.

3.3

Variable

Press the

Variable

switch to enable/disable the function.

There are two ranges of variable mV/V.

3.3.1

Coarse Control Range

The range is 0 to 500µV/V (0 to 0.5mV/V) plus the setting of the far left knob.
This range disables and replaces the center 0.1mV/V and right 0.01mV/V switches. The variable range is Switch

one plus the variable switch and will range from 0 to 0.5mV/V.

1. Press and release the

Variable

switch to enable the function.

2. Adjust the setting as required.
3. Press and release the

Variable

switch again to disable the function.

3.3.2

Fine Variable Setting

In the fine setting, the variable switch disables and replaces the far right 0.01mV/V switch. So the variable range is

switch one plus switch two plus the variable switch, it will range from 0 to 0.1mV/V.

1. Press and hold the variable switch (you must start from

Off

).

2. When the second LED (labeled 0.1mV/V) lights release the switch pushbutton.

Advertising