3.0 Load Cell Wiring

1. Route the load cell cables so they will not be damaged or cut. Cable should not be

routed near heat sources greater than 150 °F. Do not shorten any load cell cable.

The load cell is temperature compensated with the supplied length of cable. Cut-

ting the cable will affect temperature compensation. Coil and protect excess

cable so it will not be mechanically damaged or be sitting in water.

2. Provide a drip loop in all cables so that water or other liquids will not run directly

down the cables onto either the load cells or the junction box. Attach load cell

cable to the dead structure, not the vessel.

3. If conduit protection is necessary against mechanical or rodent damage to the

load cell cables, use flexible conduit and conduit adapters at the load cells.

4. Connect cables for load cells to the summing board in the junction box according

to the guide shown below and the labels on the terminal strips of the junction

box. To verify the wiring scheme, see the certification shipped with each load

cell.

5. For better performance, use positive and negative remote sense lines if the wiring

running from the junction box to the indicator is longer than 25 feet.

Drip Loop

Load Cell Wire Colors

Wire Color

Function

Negative Reading

Positive Reading

Red

+EXC

+EXC

Black

-EXC

-EXC

Green

+SIG

-SIG

White

-SIG

+SIG

Gray or Bare

Shield

Shield