... after switching on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?

The device was unable to send measured results
to the wireless receiver (seca wireless printer or
PC with seca USB wireless adapter).

– Check that the scale is integrated in the

wireless network.

– Check that the receiver is switched on.

Nearby HF equipment (e.g. mobile phones) are
interfering with reception.

- Make sure that HF equipment is kept at least

1 metre away from transmitters and receivers
in the seca wireless network.

NOTE
If such interference is not eliminated, no fur-
ther acoustic warnings are given with sub-
sequent attempts at transmission.

... only the item “SYS” is
visible in the “rF” menu?

The wireless module is disabled.

- Activate wireless module (“Activating the wire-

less module (SYS)” on page 71).

... only the “SYS” and
“Lrn” items are visible in
the “rF” menu?

The wireless module is activated but no wireless
group is set up.

- Set up wireless group (“Setting up a wireless

group (Lrn)” on page 71).

... the “APrt” and “tiME”
menu items are not visi-
ble in the “rF” menu?

No wireless printer is registered to the wireless
group.

- Use the “Lrn” menu item to log the wireless

printer into the wireless group (“Setting up a
wireless group (Lrn)” on page 71).

... when the menu is
called up, the item “rF” is
not displayed?

The wireless module of the scale is defective.

- Please notify maintenance service.

… the display “Er:X:11”
appears?

The scale has too high a load, or too great a load
is applied in one corner.

- Reduce load on scale or distribute weight

more evenly.

- Restart scale.

… the display “Er:X:12”
appears?

The scale has been switched on with too great a
load.

- Remove load.
- Restart scale.

Fault

Cause/Remedy