Physical

Characteristic

Specification

Notes

Temperature

Temperature Measurement
Range

80

F to 110 F

26.7

C to 43.3 C

Temperature Calibration
Accuracy

± 0.2

F (± 0.1 C); meets or exceeds ASTME1112-00; EN12470-3:2000

Temperature Determination
Time

predictive
(typical)

oral
axillary
rectal

4 seconds
10 seconds
15 seconds

SpO

2

Characteristic

Specification

Notes
All sensors have a measurement
range of 70% - 100%. SpO2 complies
with EN ISO 9919:2005.

Pulse Rate Range

25 to 240 bpm

Pulse Rate Accuracy

± 3 bpm
± 5 bpm

No motion
Motion

± 3 digits

Low perfusion

SpO

2

Accuracy

70-100% ± 2 digits

Adult and pediatric

70-100% ± 3 digits

Neonate

Pulse Oximetry (SpO

2

)

Saturation Range

1% to 100%

Electrosurgery interference
suppression

Yes

Sensor compatibility

For a complete list of compatible sensors refer to

www.welchallyn.com

.

Masimo

For probe/sensor compliance to EN ISO 9919:2005, see the Masimo directions for use.

Sensor lights (Masimo)

Red wavelength
Red wavelength (toe clip)
Infrared wavelength
Infrared wavelength (toe
clip)

15 mW at 50 mA pulsed

660 nm
663 nm
905 nm
880 nm

Risk mitigation: Software in this device is developed under the FDA guidance of Part 802.3 design controls and the
international standard EN 14971 risk management. These mitigate and minimize any risk associated with potential
software errors to a level as low as reasonably possible (ALARP).

Note

SpO

2

specifications are published by the SpO

2

component manufacturer and

tested by Welch Allyn using the BIO-TEK Index 2 SpO

2

simulator.

Characteristic

Specification

Dimensions

height

6.6 inches

16.8 cm

width

10.0 inches

25.4 cm

depth

6.0 inches

15.2 cm

Weight

Approximately 5.4 lbs (2.4 kg)

Color, Temperature Probe

Oral/Axillary

Blue

Rectal

Red