Blood pressure cautions, Spo2 cautions, Blood pressure cautions spo – Welch Allyn Spot Vital Signs - User Manual User Manual

Page 16: Cautions

Advertising
background image

8

Welch Allyn Spot Vital Signs

Blood Pressure Cautions

SpO

2

Cautions

Caution Extremity and blood pressure cuff motion should be minimized during
blood pressure determinations.

Caution If the blood pressure cuff is not at heart level, the difference in reading
due to the hydrostatic effect should be noted. The value of 1.80 mmHg must be
added to the displayed reading for every inch (2.5 cm) above heart level. The
value of 1.80 mmHg must be subtracted from the displayed reading for every
inch (2.5 cm) below heart level.

Caution Proper blood pressure cuff size and placement is essential to the
accuracy of the blood pressure determination. See “Chart for Determining Blood
Pressure Cuff Size” on page 23
for blood pressure cuff sizing information.

Caution The pulse oximeter is calibrated to determine the percentage of
arterial oxygen saturation of functional hemoglobin. Significant levels of
dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin may
affect the accuracy of the measurement.

Caution Physiological conditions, medical procedures, or external agents that
may interfere with the pulse oximeter’s ability to detect and display
measurements include dysfunctional hemoglobin, arterial dyes, low perfusion,
dark pigment, and externally applied coloring agents such as nail polish, dye, or
pigmented cream.

Caution Some sensors may not be appropriate for a particular patient. If at least
15 seconds of perfusion pulses cannot be observed for a given sensor, change
sensor location or sensor type for perfusion to resume.

Caution When selecting a sensor, consider the patient’s weight and activity
level, the adequacy of perfusion, the available sensor sites, the need for sterility,
and the anticipated duration of monitoring.

Advertising