Specifications, A - specifications – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 147

Advertising
background image

A

143

Specifications

ECG

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Resp

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146

SpO

2

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147

NIBP

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148

Alarms and alerts

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149

Radio

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151

Flat-panel display for the Large Color Display Interface

. . . . . . . . . . . . . . . . . . . . . . . . . 153

Environmental

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154

Physical

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154

Power

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156

Cradle

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157

ECG

Characteristic

Specification

Regulatory standards

Meets ANSI/AAMI EC13-2002 and EN 60601-1:1990 + A1 + A2

Connector

Hypertronics 9

-

pin female D01 latching connector

Selectable leads

3-lead cable

5-lead cable

Displayable

I, II or III

Internally provides

I, II or III

Displayable

I, II, III, aV

R

, aV

L

, aV

F,

V

Internally provides

I/II, III and V

Cable type detection

Automatic detection and differentiation of 3

-

or 5-lead cable

Lead fault indicator

Displayed chest icon with flashing indicator for each electrode

ECG Size (sensitivity)

0.2, 0.5, 1, 2, 4, and 8 mV/cm

Display sweep speed

25 mm/sec

Bandwidth (monitor display)

0.5 to 80 Hz

Monitor Mode

0.05 to 80 Hz

Extended Mode

Bandwidth (to Acuity)

0.05 to 80 Hz

Monitor or Extended Mode

Notch filters

50 and 100 Hz, 60 and 120 Hz, or OFF, selectable at Acuity or at the monitor

Sample rate

363.64 Hz (181.82 Hz to Acuity)

Input protection

Protected against electrosurgery and defibrillator when used with ECG cables specified
in the accessory list or at

www.welchallyn.com

.

Electrosurgery interference
suppression

Included on all vectors when Welch Allyn-approved cables are used.