Inspection and checklist report

In accordance with the maintenance interval detailed previously, the following check list
should be copied and followed.

Monitor Serial Number: _________________________________

Every six months

Inspection

Result

Checked

General examination

Visual inspection of the
monitor.

Monitor casing not broken or
cracked.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Visual inspection of the
LCD.

LCD screen undamaged.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Visual inspection of all
cable assemblies and
sensors and accessories.

Electrode cable sheathing and
connectors undamaged.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

No kinks, abrasion or wear in any
cable assembly.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

All transducers and accessories
are within their expiratory date.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Plug and socket
connectors.

Input/output connectors
undamaged.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Button check

Buttons work.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Speaker check.
Switch the monitor on by
pressing the On button.

Switch-on beeps sounded. The
standard screen is displayed.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

LED check

Mains LED on when mains
connected. Battery LED on when
battery charging.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Alarm Check

Connect SpO

2

sensor to

volunteer.

Measurement within range.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Set SpO

2

limit to lowest

setting.

Measurement out of range and
visual and audible alarm
activated.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Recurrent test

Confirm the date of last
factory inspections and
test.

If the monitor is due for a yearly
test, have a qualified technician
perform the test.

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

ˆ

Date of Inspection:

Inspector: