Leadoff, Framing Errors and DCOffset conditions:


If the application detects a Lead-off, DC offset, or Framing error within 10 seconds after you have
started an ECG recording, the recording automatically restarts.

In the case of a Leadoff condition,

the ECG recording will not start until the signal is stabilized. The progress bar of the recording will
remain at 0 until all the signals are of acceptable quality.

You can use the override button to disable the automatic restart.

DCOffSet

The DCOFFSET DETECTED! message displays in the progress bar when

battery power is interrupted or when there is a spike in voltage.


Framing Errors Framing errors display in the progress bar and can occur in CPWS

configurations when the Pro recorder is connected via a serial cable and the

packets of information are delayed.


Leadoff

The Leadoff message displays in the workspace when an electrode is

disconnected. The real-time monitor will displ

ay “Bad signal on…”. The trace

shows square waves until the signal is stabilized.