Page 20

Advertising
background image

16

Setup

Welch Allyn Propaq CS Vital Signs Monitor

Whenever you change the patient mode, the alarm limit settings, maximum NIBP cuff
inflation pressures, and internal computations are automatically changed to the
defaults for that patient mode. See page 19 for information about preset Factory
patient modes or programmable Custom patient modes.

5.

To select which vital sign waveforms will display, from the Main Menu press SETUP,
MORE, WAVE SELECT. Use NEXT and ON/OFF to turn on the desired waveforms in
the Wave Select window.

You can turn on all waveforms, but only the first four waveforms selected as ON in the
Wave Select window are displayed. You cannot turn off the ECG1 waveform.

6.

To set the HR/PR source, display sweep speed, tone volumes, and display brightness,
from the Main Menu press SETUP, MORE to access Setup Menu 2. Use NEXT and
CHANGE to select settings.

Note

If you change the patient mode, the CO

2

alarm limits in the new mode might vary

slightly from the originally-programmed CO

2

alarm limits for the new mode.

Check the CO

2

alarm limits.

CURRENT SOURCE When the selected HR/PR source is no longer available, the current

source is the active source with highest priority. The RR/BR source
cannot be manually selected. It is always CO

2

if CO

2

is active.

Otherwise, it is ECG/RESP.

SELECTED SOURCE The user-selected HR/PR source is displayed along with the HR/PR

source currently used by the monitor.

SETUP

ECG

1

: ON

WAVE SELECT

CO

2

:

ON

RESP : ON

SpO

2

: ON

NIBP : ON

ECG

2

:

ON

NEXT

ON/

OFF

INSERVICE

PREVIOUS

MENU

PA

: ON

P

2

: ON

SETUP

CURRENT SOURCE : ???

SELECTED SOURCE : ECG

SWEEP (mm/s)

ALARM TONE

HR/PR TONE

PATIENT MODE

BRIGHTNESS

HR/PR RR/BR

NEXT

CHANGE

WAVE

SELECT

MORE

???

---

:

25.0

6.25

:

:

:

:

LOW

OFF

ADULT

NORMAL

PRINTER

? ? ? is displayed
until a source is
active.

Advertising