Welch Allyn Disposable Blood Pressure Cuffs with FlexiPort Technology - User Manual User Manual

Directions for use, Introduction, Disposable blood pressure cuffs with flexiport

Advertising
background image

Disposable Blood Pressure Cuffs with FlexiPort* Technology

Directions for Use

For 30503-series and 30506-series Blood Pressure Cuffs with FlexiPort* Technology and Aneroid
Sphygmomanometers

Standards

The blood pressure cuffs are designed to function within the limits prescribed by:

• American National Standard ANSI/AAMI SP10, 2002, manual, electronic, or automated

sphygmomanometers (except for section 4.2.2a for size 13 disposable cuffs)

• Specification for noninvasive sphygmomanometers - Part 1: General Requirements: EN1060-1:1996

(except for section 9.3a for sizes 6, 7, 8, 9 disposable cuffs)

• Supplementary requirements for mechanical sphygmomanometers: EN1060-2:1996
• Supplementary requirements for electro-mechanical blood pressure measuring systems: EN 1060-3:1997

Introduction

Intended Use

Cardinal Health Blood Pressure Cuffs are noninvasive and are intended for use with manual and automated
noninvasive blood pressure measurement devices.

Contraindications

Cardinal Health

Blood Pressure Cuffs are contraindicated for neonate use. Do not use on neonate patients.

Warnings

A warning statement in this manual identifies a condition or practice that, if not corrected or discontinued
immediately, could lead to injury, illness, or death.

Cautions

A caution statement in this manual identifies information within the manual to avoid equipment failure.

CAUTION: U.S. federal law restricts the blood pressure cuffs to sale by or on the order of a
physician or licensed healthcare practitioner. This device should be used by trained
personnel.

WARNINGS: Only use the cuff when visible artery index marker falls within the range markings
indicated on the cuff, otherwise erroneous readings may result.
Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently
for irritation.
Allow space for 1 to 2 fingers between patient and cuff.
Do not apply cuff to limbs used for IV infusion.
Minimize cuff movement and limb motion during readings.
Ensure an airtight seal at all connection points prior to use.
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or air may enter patient.

CAUTIONS: Do not press cuff with a hot iron.
Do not inflate cuff unless the hook and loop is closed.
Do not allow foreign debris to ingress into tubes or port on cuff.
Do not use steam or heat to sterilize the cuff or tubing.
Do not exceed 250 mmHg with thigh disposable cuffs at or above 30° C (86° F).
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or fluid may enter the cuff.

Disposable Blood Pressure Cuffs with FlexiPort* Technology

Directions for Use

For 30503-series and 30506-series Blood Pressure Cuffs with FlexiPort* Technology and Aneroid
Sphygmomanometers

Standards

The blood pressure cuffs are designed to function within the limits prescribed by:

• American National Standard ANSI/AAMI SP10, 2002, manual, electronic, or automated

sphygmomanometers (except for section 4.2.2a for size 13 disposable cuffs)

• Specification for noninvasive sphygmomanometers - Part 1: General Requirements: EN1060-1:1996

(except for section 9.3a for sizes 6, 7, 8, 9 disposable cuffs)

• Supplementary requirements for mechanical sphygmomanometers: EN1060-2:1996
• Supplementary requirements for electro-mechanical blood pressure measuring systems: EN 1060-3:1997

Introduction

Intended Use

Cardinal Health Blood Pressure Cuffs are noninvasive and are intended for use with manual and automated
noninvasive blood pressure measurement devices.

Contraindications

Cardinal Health

Blood Pressure Cuffs are contraindicated for neonate use. Do not use on neonate patients.

Warnings

A warning statement in this manual identifies a condition or practice that, if not corrected or discontinued
immediately, could lead to injury, illness, or death.

Cautions

A caution statement in this manual identifies information within the manual to avoid equipment failure.

CAUTION: U.S. federal law restricts the blood pressure cuffs to sale by or on the order of a
physician or licensed healthcare practitioner. This device should be used by trained
personnel.

WARNINGS: Only use the cuff when visible artery index marker falls within the range markings
indicated on the cuff, otherwise erroneous readings may result.
Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently
for irritation.
Allow space for 1 to 2 fingers between patient and cuff.
Do not apply cuff to limbs used for IV infusion.
Minimize cuff movement and limb motion during readings.
Ensure an airtight seal at all connection points prior to use.
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or air may enter patient.

CAUTIONS: Do not press cuff with a hot iron.
Do not inflate cuff unless the hook and loop is closed.
Do not allow foreign debris to ingress into tubes or port on cuff.
Do not use steam or heat to sterilize the cuff or tubing.
Do not exceed 250 mmHg with thigh disposable cuffs at or above 30° C (86° F).
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or fluid may enter the cuff.

Disposable Blood Pressure Cuffs with FlexiPort*

Technology

Directions for Use

For 30503-series and 30506-series Blood Pressure Cuffs with FlexiPort* Technology and Aneroid
Sphygmomanometers

Standards

The blood pressure cuffs are designed to function within the limits prescribed by:

• American National Standard ANSI/AAMI SP10, 2002, manual, electronic, or automated

sphygmomanometers (except for section 4.2.2a for size 13 disposable cuffs)

• Specification for noninvasive sphygmomanometers - Part 1: General Requirements: EN1060-1:1996

(except for section 9.3a for sizes 6, 7, 8, 9 disposable cuffs)

• Supplementary requirements for mechanical sphygmomanometers: EN1060-2:1996
• Supplementary requirements for electro-mechanical blood pressure measuring systems: EN 1060-3:1997

Introduction

Intended Use

Cardinal Health

Blood Pressure Cuffs are noninvasive and are intended for use with manual and automated

noninvasive blood pressure measurement devices.

Contraindications

Cardinal Health

Blood Pressure Cuffs are contraindicated for neonate use. Do not use on neonate patients.

Warnings

A warning statement in this manual identifies a condition or practice that, if not corrected or discontinued
immediately, could lead to injury, illness, or death.

Cautions

A caution statement in this manual identifies information within the manual to avoid equipment failure.

CAUTION: U.S. federal law restricts the blood pressure cuffs to sale by or on the order of a
physician or licensed healthcare practitioner. This device should be used by trained
personnel.

WARNINGS: Only use the cuff when visible artery index marker falls within the range markings
indicated on the cuff, otherwise erroneous readings may result.
Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently
for irritation.
Allow space for 1 to 2 fingers between patient and cuff.
Do not apply cuff to limbs used for IV infusion.
Minimize cuff movement and limb motion during readings.
Ensure an airtight seal at all connection points prior to use.
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or air may enter patient.

CAUTIONS: Do not press cuff with a hot iron.
Do not inflate cuff unless the hook and loop is closed.
Do not allow foreign debris to ingress into tubes or port on cuff.
Do not use steam or heat to sterilize the cuff or tubing.
Do not exceed 250 mmHg with thigh disposable cuffs at or above 30° C (86° F).
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or fluid may enter the cuff.

Disposable Blood Pressure Cuffs with FlexiPort*

Technology

Directions for Use

For 30503-series and 30506-series Blood Pressure Cuffs with FlexiPort* Technology and Aneroid
Sphygmomanometers

Standards

The blood pressure cuffs are designed to function within the limits prescribed by:

• American National Standard ANSI/AAMI SP10, 2002, manual, electronic, or automated

sphygmomanometers (except for section 4.2.2a for size 13 disposable cuffs)

• Specification for noninvasive sphygmomanometers - Part 1: General Requirements: EN1060-1:1996

(except for section 9.3a for sizes 6, 7, 8, 9 disposable cuffs)

• Supplementary requirements for mechanical sphygmomanometers: EN1060-2:1996
• Supplementary requirements for electro-mechanical blood pressure measuring systems: EN 1060-3:1997

Introduction

Intended Use

Cardinal Health Blood Pressure Cuffs are noninvasive and are intended for use with manual and automated
noninvasive blood pressure measurement devices.

Contraindications

Cardinal Health Blood Pressure Cuffs are contraindicated for neonate use. Do not use on neonate patients.

Warnings

A warning statement in this manual identifies a condition or practice that, if not corrected or discontinued
immediately, could lead to injury, illness, or death.

Cautions

A caution statement in this manual identifies information within the manual to avoid equipment failure.

CAUTION: U.S. federal law restricts the blood pressure cuffs to sale by or on the order of a
physician or licensed healthcare practitioner. This device should be used by trained
personnel.

WARNINGS: Only use the cuff when visible artery index marker falls within the range markings
indicated on the cuff, otherwise erroneous readings may result.
Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently
for irritation.
Allow space for 1 to 2 fingers between patient and cuff.
Do not apply cuff to limbs used for IV infusion.
Minimize cuff movement and limb motion during readings.
Ensure an airtight seal at all connection points prior to use.
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or air may enter patient.

CAUTIONS: Do not press cuff with a hot iron.
Do not inflate cuff unless the hook and loop is closed.
Do not allow foreign debris to ingress into tubes or port on cuff.
Do not use steam or heat to sterilize the cuff or tubing.
Do not exceed 250 mmHg with thigh disposable cuffs at or above 30° C (86° F).
Intravenous Systems (IV) - Do not connect cuffs with Luer lock connectors to intravenous fluid
systems or fluid may enter the cuff.

Advertising