Page 86

Confidential

Welch Allyn OEM Technologies

Menu Options Description

The File menu contains the following items:

File\Calibration (LC-101 Module only) - Calibration is available under the File menu. Two
different calibrations are available; Zero and Two Point. The Zero calibration is a single point
calibration performed with either room air or reference air with 0% CO

2

. The Two Point

calibration requires a zero reference gas and a span reference gas. The zero can again be room
air or 0% reference gas. The span reference gas must have a CO

2

concentration of between 8%

and 12%, with the exact concentration known to within 0.1%. The span gas concentration must
be manually entered in the Configure menu item.

File\Capture - The Capture feature provide a means of logging CO

2

data and Breath Parameters

to a file. Select the Capture Setup menu item to specify the file name, data logging rate and what
data is to be logged. The data is stored as ASCII text in a CSV format in the application directory.

File\Simulate - The Simulate feature disables any communications to the CO

2

module and

displays a simulated CO

2

waveform and Breath Parameters. Clicking on this menu item enables

and disables the simulation feature.

File\Print Form - The Print Form selection prints the Main Screen to the default printer.

File\Exit - The Exit selection stops the program.