Cleaning
1. Prior to cleaning, turn power off and disconnect the power cord from both the
wall outlet and the Curon Medical Light Source.
2. Clean the external surfaces by wiping with a cloth dampened with mild soap
and water or by using the following cleaning/disinfecting solutions. DO NOT
IMMERSE.
• Cidex,
CidexPlus
•
70% isopropyl alcohol
• 10% mild bleach solution
• Metracide
• 10%
Wescodyne
• Banicide
•
Wavecide-01
3. Wipe clean the power cord by using the same cleaning/disinfecting solutions
listed above. Wipe dry with a clean cloth. Take care not to get the plug prongs
wet. DO NOT IMMERSE.
4. Do not plug the unit into a wall outlet until it is thoroughly dry.
NOTE: Do not sterilize or immerse the Curon Medical Light Source.