Avoid Electrostatic Discharge
When humidity in the working environment decreases, the human body and other
insulators can become charged with static electricity due to friction.
To prevent unwanted electrostatic discharge (ESD), follow these standard guidelines:
•
Maintain the recommended humidity of 40% to 60% in the work environment.
•
Dissipate electrostatic charge before performing routine operator maintenance.
Specifications
Monitor (Physical) Specifications
Monitor (Environmental) Specifications
Electrosurgery
Propaq Encore monitors are suitable for use in the presence of electrosurgery.
Characteristic
Specification
Recovery time following
defibrillator discharge
Less than or equal to 10 seconds
Characteristic
Specification
Degree of protection against
ingress for monitors without
CO
2
or printer options
IPX1 rating, drip proof per EN60529: 1991