Cleaning – Welch Allyn Solarc Light Source - User Manual User Manual

Page 17

Advertising
background image

Directions for Use

Chapter 3 Maintenance

15

Q-tip moistened with alcohol. Do not leave any lint on the
lamp.

6. Connect the new lamp to the connector and insert lamp to

properly seat the alignment pin located on the bottom of the
lamp into the slot located on the lamp bracket.

Figure 4. Bottom View of the Lamp Slot

7. Close the retaining latch and replace the cover. Secure the

five screws on the Solarc Light Source. Reconnect the power
supply.

Do not operate the device without the cover in place. Dispose
of the lamp according to local, state, and federal laws.

Cleaning

Clean the Solarc Light Source monthly, or as required, as follows:

1. Prior to cleaning, turn power off and disconnect the power

cord from both the wall outlet and the Solarc Light Source.

Lamp Slot.

Caution Do not submerge the S

o

larc Light Source in

water, cold steriliants, or in an ultrasonic cleaner.

Caution Do not use an autoclave, chemiclave,
glutaraldehyde, iodophor or any other sterilization method,
or damage may result.

Advertising