Propaq Encore Version 1.20

33

English

Connecting to the Acuity System

1.

If the Propaq Encore has already been connected to the patient, save the
patient’s Trends and Alarm Limit settings by keeping the monitor turned
on. (The Propaq Encore transmits up to five hours of trend information
when you connect the Propaq Encore to the Acuity network.)

If the Propaq Encore has not been connected to the patient, clear any prior
patient’s trends and alarm limit settings by switching off the Propaq
Encore and after a moment, switching it on again.

2.

If the Propaq Encore is not already connected to the patient, attach leads
and sensors to the patient as directed by the Propaq Encore Reference Guide.

3.

Plug in the Acuity network cable to the Acuity network jack on the Propaq
Encore side panel shown in the following figure. Plug in the other end of
the cable to the bedside Acuity network jack.

Right Side Panel

WARNING

Connect the Propaq Encore to an Acuity system only. Connecting to other
networks (such as telephone systems that use the same type of connector)
could damage the monitor or injure the patient. If in doubt about the network
jacks or devices, consult your facility’s Biomedical Engineering Department.

DEFIB SYNCHRO

MONITOR

EKG x 1000

640-0326-00 Rev.2

!

3A

2AG

12-28V, 3A

!

!

!

Acuity Connector

Power Connector