Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

8

9

Missing probe well.

Probe (well)/

Equipment alert

63

E4.9

Insert probe

well.

Indicate

missing

probe well

8

10 Problem saving to the

module's EEPROM in

biotech mode.

Module/ Host

controller alert

65

E4.3

Verify that

batteries are

fresh. Try to

make changes

again. If the

problem

persists,

replace the

module.

Indicate

internal

error

8

11 The module's error

detection mechanism

detected an error.

Module/

Equipment alert

66

E0.8

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

12 The module's probe

error detection

mechanism detected an

error.

Probe/ Equipment

alert

66

E0.8

Probe error,

try again. If

the problem

persists,

replace the

probe.

Indicate

probe

error

8

13 The module's log error

detection mechanism

detected an error.

Module/

Equipment alert

Internal error

log problem;

error log will

be cleared.

Restore

default

error log

8

14 The module's calibration

error detection

mechanism detected an

error.

Module/

Equipment alert

Try again. If

the problem

persists,

replace the

module.

Indicate

module

error

8

15 No probe connected.

Probe/ Equipment

alert

Connect a

probe and try

again. If a

probe is

connected,

replace the

probe. If the
problem

persists

replace the

module.

Indicate

probe

error