About the connectivity software modules

The Connex CSK software comprises a set of modules that are installed separately as
needed. The connectivity modules are listed here:

Firewall settings

If you are using a firewall between any PCs in your configuration, the following ports or
application files are required to gain access through the firewall. Software installation
automatically opens these ports.

Module

Full name

Description

Installation location

CPSAPI

CardioPerfect
Server
Application
Programming
Interface

This is the database connectivity module. It plays a
central role, directing communications between
various software modules and the server software.

One CPSAPI per site

DC

Data Catcher

This module receives data from the device(s).

One DC per CPSAPI installation
and per Workstation that supports
wired-USB

DCP

Device
Communication
Protocol

This module listens for device(s) that are
connected wirelessly, and it tells them which IP
address to communicate with.

One DCP per subnet

FI

File Importer

This module receives tests from removable media
(SD memory cards or USB storage devices).

One FI per Workstation receiving
tests via removable media

Software module

Port

Type

Application file

CPSAPI

9247

TCP

CPWS.Service.exe

DCP

44435

UDP

DCPd.exe

DCP

7711

UDP

DCPd.exe

DC

281

TCP

DataCatcher.exe

CP 200

ICMP

Allow incoming echo requests

Ping test