US-MINIBENCH Use & Care Instructions

Thank you for purchasing a US Balance pocket scale.

We are confident that with normal use and proper treatment your scale will provide you with years of

reliable service. Please read all the operation instructions carefully before you use your new scale.

Display Messages:

LO — Low battery; please change the batteries in the scale.

OuEr — The load cell of the scale is being overloaded. You need to remove the excessive

weight from the scale immediately.

UnSt — The scale is placed on an unstable surface, please move the scale to a more stable

surface.

rCAL — the scale cannot return to zero, please recalibrate the scale.

CAL — This is the prompt for beginning the calibration process of your scale. You will see

this message when you press and hold the Cal button on your scale. Once this message appears,

you must release the button and follow the remaining steps for calibrating your scale.

PASS — you have successfully calibrated your scale, you are now able to use the scale.

FAIL — Calibration has failed, please make sure you have the scale on a stable surface and that
you are using the correct weight to calibrate the scale. Try to calibrate the scale following the
calibration instructions exactly.

Advertising