Sample Preparation Procedure for Crude Fat –

for HIGH fat samples

To prepare meat samples with > 15% fat

by weight or plant samples with > 20% fat by weight for fat extraction,

follow the procedure detailed below.

1. Using a Solvent Resistant Marker, number all of the weigh tins and empty

filter bags you will use during the fat extraction procedure. Use one weigh tin
per filter bag. Although it is not necessary in the determination of fat, it is
recommended that you weigh and record the weights of all the filter bags
prior to filling them in order to calculate the moisture value for the samples.

% Moisture = 100 x ((Filter Bag Weight + Sample Weight) - (Weight after drying - Tin Weight))

Sample Weight

2. Dry all weigh tins in an oven at 102°C±2° for 15 minutes.

3. Remove weigh tins from the oven and place them in a

Weigh Tin Desiccator or in a Desiccant Pouch.

4. Allow weigh tins to cool to room temperature.

5. Weigh and record tin weights (W

4

).

6. Place an empty filter bag in the Bag Holder in an open position.

7. Tare the weight of the empty filter bag and the holder together.

8. Add 1.5 – 2g of sample to the filter bag. Keep all particles away from the

sealing area of the filter bag.

9. Record the weight of the sample (W

1

) and tare.

10. Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)