Table III-3: Accuracy for Low Torque CAP 1000+

Cone

≤400 RPM

70 RPM

900 RPM

900 RPM

10-100% FSR

10-100% FSR

50% FSR

>50% FSR

01

±2.0%

±2.0%

±2.0%

±2.0%

02

±2.0%

±2.0%

±2.0%

±2.0%

03

±2.0%

±2.0%

±2.0%

±2.0%

04

±2.0%

±2.0%

±2.0%

±2.0%

05

±2.0%

±2.0%

±2.0%

±2.0%

06

±2.0%

±2.0%

±2.0%

±2.0%

07

±2.0%

*

*

*

08

±2.0%

±2.0%

±2.0%

±2.0%

09

±2.0%

±2.0%

±2.0%

±2.0%

10

±2.0%

±2.0%

±2.0%

±2.0%


* Not recommended for use at these RPM

Table III-4: Accuracy for High Torque CAP 1000+

Cone

≤400 RPM

70 RPM

900 RPM

900 RPM

10-100% FSR

10-100% FSR

50% FSR

>50% FSR

01

±2.0%

±2.0%

±2.0%

±4.0%

02

±2.0%

±2.0%

±2.0%

±4.0%

03

±2.0%

±2.0%

±2.0%

±4.0%

04

±2.0%

±3.0%

±3.0%

±6.0%

05

±2.0%

±4.0%

±4.0%

±8.0%

06

±2.0%

±5.0%

±5.0%

±10.0%

07

±2.0%

*

*

*

08

±2.0%

*

*

*

09

±2.0%

*

*

*

10

±2.0%

±2.0%

±2.0%

±2.0%


* Not recommended for use at these RPM

The specification of temperature accuracy on CAP Viscometers is stated below:

LOW TEMP UNITS:
• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point

on the viscometer is 5°C to 75°C.

• In ambient conditions outside of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature

set point on the viscometer is 5°C to 75°C.

HIGH TEMP UNITS:
• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point

on the viscometer is 50°C to ≤100°C.

• In ambient conditions of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature set point

on the viscometer is 101°C to 235°C.