Brookfield Dial Viscometer User Manual

Page 11

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 11

Manual No. M/85-150-P700

Turn the adjustment ring to the right (counterclockwise as you look down on the instrument) in small
increments (one or two divisions of the ring) while watching the Viscometer dial (Figure A5).
Remember that you are trying to raise the cup so that the pin in the cone touches the pin in the cup. Once
you have found this hit point, you can "back off" to create the desired gap between cone and cup. You
must wait at least 6 seconds between movements of the ring.

Turn the adjustment ring until the display reading jumps from 0.0 to 0.5% (HA and HB models) or from
0.0 to 0.5 % (RV models) or 0.0 to 1.0% (LV models). This is the HIT POINT. The cup may have to
be adjusted up and down several times before this deflection is satisfactorily reached. The LV series
Viscometer will be the most sensitive.

Figure A5

Turn Adjusting Ring clockwise until

the % torque display is stable at 0.0%

Adjustment Ring

Index Mark

Turn Adjustment Ring Counter-

clockwise in Small Increments

Adjustment Ring

Index Mark

Figure A4