Brookfield DV2+Pro Viscometer User Manual

Page 38

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 38

Manual No. M03-165-F0612

print mode and finally initiate either of the timed modes of operation. While setting

up the timed mode parameters, continuous print operation will cease. However, upon

starting the timed operation, the DV-II+Pro will output an initial data string to the printer

and then continue printing data strings (at the user defined time interval) for the duration

of the timed run. At the end of the timed run, continuous printing will again be disabled

and the user may print single strings (of the final data point) at your option until you exit

the timed mode. Upon returning to the default operation mode, continuous printing will

again resume at the user selected time interval. In a similar manner, if you are in the

once-per-

PRINT

-key-press mode, when you enter the timed mode of operation you will

be able to print data strings at any time during the timed mode by pressing the

PRINT

key.

IV.3.3 Print Time Interval

This option is used to set the print time interval to the selected printer. Scroll to Set Print Time,

as shown in Figure IV-29:

D TIMED TO TORQUE
SET PRINT TIME

Figure IV-29

Press the ENTER/AUTORANGE key. On entry, Figure IV-30 is displayed:

PRINT INTERVAL
SET MIN’S: 00

↑↓

Figure IV-30

Using the

UP

and

DOWN ARROW

keys, enter a value for the minutes between successive print strings.

This value can be as high as fifty-nine (59) minutes and as low as 00.

When satisfied, press the

OPTIONS/TAB

key to enter the seconds setting display:

PRINT INTERVAL
SET SEC’S: 00

↑↓

Figure IV-31

Using the

UP

and

DOWN ARROW

keys, enter a value for the seconds portion of the print interval.

This value can be between zero (0) and fifty-nine (59) seconds.

Note:

The value for minutes or seconds must be other than zero (0) or

you will print continuously when you exit this mode. A press of

the MOTOR ON/OFF/ESCAPE key would exit this option and

take you back to the screen of Figure IV-29.

Continued pressing of the

OPTIONS/TAB

key will toggle between the minutes and seconds input

screens. Press the

ENTER/AUTORANGE

key to accept the new values for print interval in minutes and

seconds. You will now be in the screen display of Figure IV-29 where you may re-enter the print

interval mode, or exit to the default screen (Figure II-5) by pressing the

MOTOR ON/OFF/ESCAPE

key.

#
$

Advertising