Brookfield DV3T Rheometer User Manual

Page 108

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 108

Manual No. M13-2100

Calibration Procedure for a Small Sample Adapter

Brookfield recommends a two step check. First, verify the calibration of the rheometer using the

standard rheometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as

detailed in this appendix. Second, verify the calibration of the rheometer using the Small Sample

Adapter. The use of an accessory device may increase the accuracy of measurement associated

with the DV3T.

When a Small Sample Adapter is used, the water jacket is connected to the water bath and the

water is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount

varies with each spindle/chamber combination (refer to the Small Sample Adapter

instruction manual).

2) Place the sample chamber into the water jacket.
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free

hanging spindle (or directly attach the solid shaft spindle) to the DV3T.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test

temperature.

5) Measure the viscosity and record the Rheometer reading.

Note: The spindle must rotate

at least five (5) times before readings are taken.

Calibration Procedure for a Thermosel System

Brookfield recommends a two step check. First verify the calibration of the rheometer using the

standard rheometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as

detailed in this appendix. Second verify the calibration of the rheometer using the Thermosel. The

use of an accessory device may increase the accuracy of measurement associated with the DV3T.

When a Thermosel System is used, the controller stabilizes the Thermo Container at the test

temperature.

1) Put the proper amount of HT viscosity standard fluid into the HT-2 sample chamber.

The amount varies with the spindle used (refer to the Thermosel instruction manual).

2) Place the sample chamber into the Thermo Container.
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free

hanging spindle (or directly attach the solid shaft spindle) to the DV3T.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test

temperature.

5) Measure the viscosity and record the Rheometer reading.

Note: The spindle must rotate

at least five (5) times before readings are taken.

Calibration Procedure using UL or DIN Adapters

Brookfield recommends a two step check. First, verify the calibration of the rheometer using the

standard rheometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as

detailed in this appendix. Second, verify the calibration of the rheometer using the UL or DIN

Adapters. The use of an accessory device may increase the accuracy of measurement associated

with the DV3T.

Advertising