Appendix E - Calibration Check Procedures

For more help go to

www.brookfieldengineering.com and download the video.

Brookfield’s accuracy statement for viscometers used with standard spindles is +/-1% of Full Scale

Range. When measuring viscosity with a specific spindle rotating at a defined speed, the maximum

viscosity that can be measured is defined as Full Scale Range. For digital viscometers this value is

easily determined by pressing the “AUTORANGE” key. The display will show the Full Scale Range

viscosity in cP or mPa.s and the torque value will show 100%. Multiply the Full Scale Range viscosity

by 1% to determine the accuracy of any future measurement made with that spindle/speed combination.

When using the following accessory devices with your viscometer, the accuracy is +/-2%. Dimensional

tolerances in the accessory device allow for increase from +/-1% to +/-2%.

• Small Sample Adapter

• Thermosel

• UL Adapter

• DIN Adapter

• Spiral Adapter

The accuracy of the DV-I PRIME is verified using viscosity standard fluids, which are available from

Brookfield Engineering Laboratories or your local Brookfield agent. Viscosity standards are Newtonian,

and therefore, have the same viscosity regardless of spindle speed (or shear rate). Viscosity standards,

calibrated at 25°C, are shown in Table E-1 (Silicone Oils) and Table E-2 (Mineral Oils).

Container size

: For Viscosity Standards < 30,000 cP, use a 600 mL Low Form Griffin Beaker

having a working volume of 500 mL.
Inside Diameter: 3.25”(8.25cm)

Height: 4.75”(12.1cm)
For Viscosity Standards

30,000 cP, use the fluid container.

Note

: Container may be larger, but may not be smaller.

Temperature

:

As stated on the fluid standard label: (

+

/-) 0.1

°

C

Conditions

:

The DV-I

PRIME should be set according to the operating instructions. The

water bath must be stabilized at test temperature. Viscometers with the letters

LV” or “RV” in the model designation must have the guard leg attached,

see Appendix F for more information on the guard leg).

Normal 25ºC Standard Fluids

Viscosity (cP)

Viscosity (cP)

High Temperature Standard Fluids

Three Viscosity/Temperatures**

5

5,000

HT-30,000

10

12,500

HT-60,000

50

30,000

HT-100,000

100

60,000

500

100,000

**25ºC, 93.3ºC, 149ºC

1,000

Refer to Brookfield catalog for more information

Table E-1