Table D1 (continued)

Table D2 lists the model codes and spring torque constants for each Rheometer model.

The full scale viscosity range for any DV-III model and spindle may be calculated using the equation:

Where:

TK

=

DV-III Torque Constant from Table D2

SMC

=

Spindle Multiplier Constant listed in Table D1

SC4-27

27

25

0.34

SC4-28

28

50

0.28

SC4-29

29

100

0.25

SC4-31

31

32

0.34

SC4-34

34

64

0.28

SC4-37

37

25

0.36

CP40

40

0.327

7.5

CP41

41

1.228

2

CP42

42

0.64

3.8

CP51

51

5.12

3.84

CP52

52

9.83

2

Spiral Adapter

70

105

0.677

Thermosel DIN spindle

81

3.7

1.29

SSA DIN spindle

for 13R or 13RP chamber

82

3.75

1.29

SSA DIN spindle

for 7R or 7RP chamber

83

12.09

1.29

ULA DIN spindle

85

1.22

1.29

ULA DIN spindle

86

3.65

1.29

ULA DIN spindle

87

12.13

1.29

SPINDLE

ENTRY CODE

SMC

SRC

LVDV-III

0.09373

LV

2.5xLVDV-III

0.2343

4L

5xLVDV-III

0.4686

5L

1/4

RVDV-III

0.25

1R

1/2

RVDV-III

0.5

2R

RVDV-III

1

RV

HADV-III

2

HA

2xHADV-III

4

3A

2.5xHADV-III

5

4A

HBDV-III

8

HB

2xHBDV-III

16

3B

2.5xHBDV-III

20

4B

5xHBDV-III

40

5B

Table D2

MODEL

TK

MODEL CODE

ON DV-III SCREEN

Full Scale Viscosity Range [cP] = TK SMC

10,000
RPM

*

*

Advertising