II. OPERATION

II.1 Set-Up

1) Be sure that the power switch is off. Attach the power cord to the appropriate power source.

2) Move the operating handle to the top (upper most) position.

3) Connect the printer (if used) to the parallel port. Be sure the printer power is off.

4) Loosen the thumb screw on the Viscometer shaft; insert the paddle spindle into the Viscometer

shaft, as far as it will go. Line up the groove on the spindle with the hole that the thumb screw
goes into. Tighten the thumb screw.

5) For Quart Cans: The can will be placed directly on the viscometer base.

6) For Pint Cans: Place the pint can adapter (KU1-74) on the viscometer base, pulling the spring

loaded front locator out and placing the adapter against the rear locating pins. Release the
locator to secure the adapter in place.

7) For 1/2 Pint Cans: Follow the procedure for pint cans. Place the 1/2 pint

can adapter (KU1-73) on top of the pint can adapter (KU1-74).

II.2 Taking Measurements

1) Pour sample into the container. Fill to 3/4 inch (20 mm) of the container

lip.

2) Bring the sample in the container to the specified temperature.

3) Switch on the KU-1+ power switch and then the printer (if used). Switch

the Hold Reading Switch to the up position. (See
Figure 5.)

4) Select the desired display (KU or gm) using the

Display unit Selector switch.

5) For Quart Cans: Place the sample container on

the Viscometer base by pulling the spring loaded
front locator out, placing the container against the
rear of the locating pins, and releasing the locator
to secure the can in place.

6) For Pint & 1/2 Pint Cans: Place the sample

container directly on the can adapter mounted on
the viscometer base.

7) Move the Viscometer handle down to the lowest

position. This will automatically immerse the
spindle into the fluid. If the correct amount of fluid

KU-1+

VISCOMETER

POWER

KREBS UNITS

gm

HOLD

Figure 5

Figure 6

KU1-74
(pint adapter)

KU1-73
(1/2 pint adapter)

Knob