Brookfield QTS25 Texture Analyzer User Manual

Page 14

Advertising
background image

QTS MATERIALS EVALUATION SYSTEM OPERATOR MANUAL

Page 14

Select utility : < .. Monitor load .. >


b) < .. Cal. distance .. >

Select utility: Cal. distance .. >

Press 'ENTER' Set calibrated probe

just above platen

Screw in 80mm calibration probe and use the up and down

↑↓ keys to position

probe approximately 3mm above platen.

Upon selection of this option the 'CALIBRATE' LED on QTS25 straining frame will

be lit

Press 'ENTER'

Nulling Cell

Beam

Creeping

When the probe touches the platen the beam stops.

Entered Calibrated Length

Key in 80 and press 'ENTER'

New

Distance

Calibration

Set

The beam moves upwards to allow the calibration probe to be removed and the

display reverts to :

QTS Controller READY

"Time" "Day" "Date"

c) < .. Exit from menu .. >

Select utility : < .. Exit from menu

Press 'ENTER' QTS Controller READY

"Time" "Day" "Date"

d) Statistics Functions .. >

Select utility: < Statistics Functions >

Press

'ENTER'

Select

option:

• Invalidate Result

• Recover Result

• Erase All Statistics

• Exit from Menu