Iii. specifications for test set up – Brookfield Spiral Adapter User Manual

Page 7

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 7

Manual No. M93-270-D0812

III. SPECIFICATIONS FOR TEST SET UP

The Spiral Adapter can be used with any Brookfield Viscometer/Rheometer. Selection of the appropriate

Viscometer/Rheometer depends upon the viscosity range of the sample fluid over the shear rates at

which it will be tested.
There are four basic spring torque series offered by Brookfield as shown in Table 1. The higher the

torque range, the higher the viscosity measurement range. The corresponding ranges for viscosity

measurement when the spiral spindle is rotating at 1.0 RPM are also shown. Lower spindle speeds are

possible, but the pumping action within the spiral chamber may not be sufficient to provide meaningful

viscosity data.

Brookfield

Viscometer/Rheometer

Series Designation

Spring Torque

Spiral Adapter

Viscosity Range

at 1.0 RPM

cP (mPa·s)

dyne-cm

milli Newton-m

minimum

maximum

LV

673.7

0.0673

1.0 x 10

4

1.0 x 10

5

RV

7,187.0

0.7187

1.1 x 10

5

1.1 x 10

6

HA

14,374.0

1.4374

2.2 x 10

5

2.2 x 10

6

HB

57,496.0

5.7496

9.0 x 10

5

9.0 x 10

6

Table 1

Table 2

provides general information on Brookfield Viscometers/Rheometers and the type of data that

can be generated when using the Spiral Adapter.
Measurement accuracy for any viscosity value obtained during a test is ± 2.0% of the full scale viscosity

range (i.e. at 100% torque) for the specific rotational speed in use. See Appendix A for the necessary

information to determine full scale range.

Model

Dial

DV-E

DV-I Series

DV-II Series

DV-III Series

Speeds

8

18

18

54

2,600

Shear Rate

0.677 sec

-1

per RPM

Signal Options

Torque

Viscosity

Temperature

No
No
No

No
No
No

Yes

No

*

Yes

Yes

Yes

Yes

Yes

Yes

Software

No

No

No

Yes

Yes

Temperature

Probe

No

*

Yes

Table 2

Note: At speeds of 1 RPM and lower, additional time may be required to allow for

complete deflection of the torque sensor.

* Available as option when ordering instrument.

Advertising