I.4 Operation

Viscometer and Stand

The instrumentation should be placed on a firm level surface near a 15 amp, 115 or 230 volt, 50 or

60 Hz grounded A.C. electrical service, as required. A cooling water supply and drain ar rangement

is also recommended for cooling the thermo-container with optional cooling plug assembly.

Set up the viscometer stand by connecting the upright rod with clamp attached into the base. Screw

the leveling screws into position on the base. (Refer to the assembly instruction supplied with your

lab stand.) Attach the vis cometer to the stand by inserting it into the clamp; level the viscometer

and position it centered between the stand legs. Raise the viscometer on the lab stand upright rod

so that there will be sufficient room to attach the alignment bracket. Check that the power switch

is OFF and plug the viscometer power cord into the appropriate electrical service.

Alignment Bracket

Attach the alignment bracket to the rear of the viscometer pivot cup, securing it tight with the

knurled screw. (Refer to Figure I-2).

Thermo-Container, Sample Chamber, Safety Guard

Using the three (3) leveling screws in the thermo-container base, to level the base.

Check that the perforated safety guard is flush onto the base of the thermo-container.

Using the extracting tool, insert the chamber into the thermo-container. Rotate the chamber until

it drops and locks in place prohibiting further rotation.

Controller and Probe

Set the controller on a level surface adjacent to the thermo-container. Insert the male plug (three-

prong) from the thermo-container braided cord into the socket on the back of the controller (turn

and lock connection). Insert the four-inch stainless steel probe (Resistance Thermometer) into the

hole in the thermo-container located directly above the braided cord. Plug the other end of the probe

into the connector located on the back of the controller. To remove the probe, depress the spring

clip and carefully slide probe out. This removal procedure should not be followed if the thermo-

container is at a temperature above 100°F (37°C). Cool the thermo-container first.

If the probe is not inserted in the thermo-container and the

controller is turned on, the thermo-container will cease to heat

and the controller will give an audible signal!

Set the controller power switch to the OFF position.

Plug the controller power cord into a 115V, 50/60 Hz or 230 V, 50/60 Hz grounded A.C. power

supply, as required.

System Alignment & Spindle Insertion

Looking down on the viscometer, level the stand base by adjusting the leveling screws until the

bubble is centered.